Technical specification
|
Frequency
|
|
Frequency Range
|
10MHz to 6000MHz
|
Setting Resolution
|
10Hz by direct keyboard entry, or in user-set increments of 10Hz to 999?999MHz by rotary control or increment–decrement keys
|
Phase Noise
|
500MHz Carrier: <-110dBc/Hz @ 20kHz offset; <-120dBc/Hz @ 100kHz
|
|
3GHz Carrier: <-95dBc/Hz @ 20kHz offset; <-110dBc/Hz @ 100kHz offset
|
|
6GHz Carrier: <-89dBc/Hz @ 20kHz offset; <-104dBc/Hz @ 100kHz offset
|
Residual FM
|
Equivalent peak deviation for 300Hz to 3.4kHz B/W: 12Hz at 500MHz carrier
|
Settling Time
|
<8ms to settle within 100Hz or 0.1ppm of final frequency, if greater
|
Reference Frequency
|
Internal or External (via rear panel BNC)
|
Internal Accuracy
|
+/- 1 ppm over 15°C to 30°C; +/- 2 ppm over 5°C to 40°C
|
Internal Stability
|
< +/-1ppm/year ageing
|
Internal Ref. Out
|
10MHz from 50 Ohms, amplitude 2V pk-pk into 50 Ohms
|
External Ref In
|
10MHz into 50 Ohms, amplitude 2V pk-pk to 5V pk-pk
|
Output Level
|
|
Level Range
|
–110dBm to +7dBm (0.1μV to 500mV into 50 Ohms)
|
Setting Resolution
|
0.1dB (or 0.01μV to 1mV) by direct keyboard entry, or in user-set increments of 0.1dB to 100dB (or 0.01μV to 100mV)
|
Accuracy
|
Better than ± 2dBm
|
Harmonics
|
<-25dBc at +7dBm. <-30dBc@levels <=0dBm 30 to 6000MHz, <-25dBc@levels <=0dBm 10 to 30MHz
|
Sub-Harmonics
|
<3000MHz : None. >3000MHz :<–40dBc (typ) @ +7dBm
|
Non-Harmonic Spurii
|
<–50dBc >10kHz offset 10 to 3000MHz, <-44dBc >10kHz offset 3000MHz to 6000MHz
|
Output Connector
|
Type N
|
Sweep
|
Frequency and amplitude sweep
|
Step Sweep
|
Step frequency and/or amplitude according to a formula over a specified number of points
|
Dwell Time
|
Individually programmable 0.01 to 10.000sec per step
|
Max Points
|
1000
|
Sweep Run
|
Continuous or single. Sweep up or down
|
Step Spacing
|
Linear or logarithmic
|
Sweep Triggering
|
Manual, Ext. signal, timed (0.1 – 999.9sec) or via remote interface
|
Sync Signal
|
Output Stable) available during dwell time. Programmable as high or low
|
List Sweep
|
As for Step Sweep except that a user defined table of frequency, amplitude and dwell time values defines the steps. The table can be created within the instrument or downloaded via the remote interfaces
|
Max Points
|
1000
|
List Storage
|
Up to 16 Sweep Lists can be stored permanently within the instrument
|
Point Trigger
|
Each point in a sweep (step or list) can be subject to a trigger event rather than a dwell time. Trigger by Manual, Ext. signal or via remote interface
|
Trim
|
(User Level Compensation Tables)
|
Function
|
A table of frequency/gain pairs allows the user to modify the generator output level with respect to frequency to calibrate an entire test set up or improve the calibration of the generator alone
|
Max Points
|
100
|
Interfaces
|
Full digital remote control through the RS232, USB, LAN and GPIB interfaces
|
RS-232
|
9-pin D connector. Variable Baud rate 1200 and 115200 Baud
|
USB
|
USB 2.0 connection. Operates as a virtual COM port
|
GPIB (IEEE-488)
|
Conforming with IEEE488.1 and IEEE488.2
|
LAN (Ethernet)
|
Standard 10/100 base-T hardware connection. ICMP and TCP/IP Protocol
|
General
|
|
Size and Weight
|
2U (86.5mm) height; half-rack (213mm) width; 350mm long. 3.65 kg. (8 lb)
|
Power
|
110-240VAC ±10% 50/60Hz; 100-120VAC ±10% 400Hz; 60VA max
|
Approvals
|
Complies with EN 61010-1 (Safety) and EN61326 (EMC)
|